RIGOL DM3058 User Manual
Page 68

RIGOL
Programming Guide for DM3058/DM3058E
3-43
Return Format:
The query returns the measured capacitance value in scientific notation, such as:
8.889030e-05, the unit is F.
34. :MEASure:CAPacitance
Command Format:
:MEASure:CAPacitance {
Function:
The command sets the capacitance range.
Parameter:
Name
Type
Range
Default
Discrete
{0|1|2|3|4|5}
2
Explanations:
For the parameter range, see table below.
Table 3-9 Capacitance range
Parameter
Range
0
2 nF
1
20 nF
2
200 nF
3
2 μF
4
200 μF
5
10000 μF
MIN
2 nF
MAX
10000 μF
DEF
200 nF
Example:
Sets the capacitance range to 2 μF.
:MEASure:CAPacitance 3
35. :MEASure:CAPacitance:RANGe?
Command Format: