RIGOL

Programming Guide for DM3058/DM3058E

4-11

1. CONFigure:VOLTage:DC

Command Format:
CONFigure:VOLTage:DC
{|MIN|MAX|DEF},{|MIN|MAX|DEF}

Function:
The command presets the multimeter as specified range and resolution for DC
voltage measurement, but does not start the measurement.

Parameter:

Name

Type

Range

Default

Discrete

{200 mV|2 V|20 V

|200 V|1000 V|AUTO}

AUTO

Discrete

Refer to Table 4-2

1ppm x range


Explanations:
The unit is V.
For : MIN = 0. 2 (200 mV), MAX = 1000, DEF=AUTO.
relates to both the measurement rate and integration time, see
table below.

Table 4-2 Resolution ranges

Resolution

Measurement

rate

Integration time

100ppm x range

10ppm x range

3ppm x range
1ppm x range

0.3ppm x range

F

M
M

S
S

0.02 NPLC*

0.2 NPLC

1 NPLC

10 NPLC

100 NPLC

*Note: 1PLC is equal to 0.02s.


2. CONFigure:VOLTage:AC

Command Format:
CONFigure:VOLTage:AC
{|MIN|MAX|DEF},{|MIN|MAX|DEF}

Function: