Measure:period – RIGOL DM3058 User Manual
Page 65

RIGOL
Programming Guide for DM3058/DM3058E
3-40
Return Format:
The query returns: 0, 1, 2, 3 or 4 (see Table 3-5).
27. :MEASure:PERiod?
Command Format:
:MEASure:PERiod?
Function:
The command queries the measured period value.
Explanations:
The command is unavailable in double-display.
Return Format:
The command queries the measured period value of AC signal in scientific
notation, such as: 9.18543e-05, the unit is s.
28. :MEASure:PERiod
Command Format:
:MEASure:PERiod {
Function:
The command sets the voltage range under period measurement.
Parameter:
Name
Type
Range
Default
Discrete
{0|1|2|3|4}
2
Explanations:
The available period range is 1μs~50ms.
For the parameter range, refer to Table 3-5.
Example:
Sets the voltage range to 20V under period measurement.
:MEASure:PERiod DEF or :MEASure:PERiod 2