Sense:]fresistance:resolution, Sense:]voltage:dc:nplc – RIGOL DM3058 User Manual
Page 138

RIGOL
Programming Guide for DM3058/DM3058E
4-45
setting of resistance.
Return Format:
The query returns the current precision of 2-wire resistance measurement.
23. [SENSe:]FRESistance:RESolution
Command Format:
[SENSe:]FRESistance:RESolution {
[SENSe:]FRESistance:RESolution? [MINimum|MAXimum]
Function:
The commands set and query the precision of 4-wire resistance measurement.
Parameter:
Name
Type
Range
Default
Refer to Table 4-2
3ppm x range
Explanations:
[SENSe:]FRESistance:RESolution {
desired precision of 4-wire resistance measurement
[SENSe:]FRESistance:RESolution? [MINimum|MAXimum] – Queries the current
precision of 4-wire resistance measurement.
4-2.
Both 2-wire and 4-wire resistance measurements are influenced by this command.
All the commands of 2-wire and 4-wire resistance will have influence over the
setting of resistance.
Return Format:
The query returns current precision of 4-wire resistance measurement.
24. [SENSe:]VOLTage:DC:NPLC
Command Format:
[SENSe:]VOLTage:DC:NPLC {0.02|0.2|1|10|100|MINimum|MAXimum}
[SENSe:]VOLTage:DC:NPLC? [MINimum|MAXimum]