Sense:]current:dc:range, Sense:]current:ac:range – RIGOL DM3058 User Manual
Page 126

RIGOL
Programming Guide for DM3058/DM3058E
4-33
[SENSe:]VOLTage:AC:RANGe? [MINimum|MAXimum] – Queries the range of AC
voltage measurement.
Return Format:
The query returns the range of AC voltage measurement.
4. [SENSe:]CURRent:DC:RANGe
Command Format:
[SENSe:]CURRent:DC:RANGe {
[SENSe:]CURRent:DC:RANGe? [MINimum|MAXimum]
Function:
The commands set and query the range of DC current measurement
Parameter:
Name
Type
Range
Default
Discrete
{ 200 µA|2 mA|20 mA
|200 mA|2 A|10 A }
200 mA
Explanations:
[SENSe:]CURRent:DC:RANGe {
range of DC current measurement from {200 µA|2 mA|20 mA|200 mA|1 A|10 A};
thereinto, MIN = 200 µA, MAX = 10 A and the unit is A. If a 200 mA range is
required, enter 0.2 as its parameter.
[SENSe:]CURRent:DC:RANGe? [MINimum|MAXimum] – Queries the range of DC
current measurement.
Return Format:
The query returns the range of DC current measurement.
5. [SENSe:]CURRent:AC:RANGe
Command Format:
[SENSe:]CURRent:AC:RANGe {
[SENSe:]CURRent:AC:RANGe? [MINimum|MAXimum]