Rate:voltage:dc, Rate:voltage:ac – RIGOL DM3058 User Manual
Page 71

RIGOL
Programming Guide for DM3058/DM3058E
3-46
1. :RATE:VOLTage:DC
Command Format:
:RATE:VOLTage:DC?
:RATE:VOLTage:DC {F|M|S}
Function:
The commands query and set the rate of DC voltage measurement.
Explanations:
:RATE:VOLTage:DC? - Queries the current measuring rate of DC voltage. The
command is valid only when DCV function is enabled.
:RATE:VOLTage:DC {F|M|S} – Sets the desired measuring rate of DC voltage. For
the range and the explanation of each parameter, see table below.
Table 3-10 The parameter range and explanations
Parameter
Explanation
Mark
Rate
Refresh rate
F (FAST)
Fast measure
F
123 reading/s
50 Hz
M (MEDIUM) Medium measure
M
20 reading/s
20 Hz
S (SLOW)
Slow measure
S
2.5 reading/s
2.5 Hz
Return Format:
The query returns the parameter related to current rate (see Table 3-10).
Example:
Sets the rate under DC voltage measurement to Medium.
:RATE:VOLTage:DC M
2. :RATE:VOLTage:AC
Command Format:
:RATE:VOLTage:AC?
:RATE:VOLTage:AC {F|M|S}
Function:
The commands query and set the rate of AC voltage measurement.
Explanations: