Calculate:dbm:state, Calculate:pf – RIGOL DM3058 User Manual
Page 35

RIGOL
Programming Guide for DM3058/DM3058E
3-10
:CALCulate:DBM:REFErence {
reference resistance of dBm.
The command is valid only when dBm operation is enabled.
The reference resistance should be an integer within 2 and 8000, the unit is Ω.
MIN is equal to 2 Ω, MAX is equal to 8000 Ω and DEF is 600 Ω.
Return Format:
The query returns the current reference resistance of dBm.
14. :CALCulate:DBM:STATe
Command Format:
:CALCulate:DBM:STATe?
:CALCulate:DBM:STATe {ON|OFF|1|0}
Function:
The commands query and set the dBm state.
Explanations:
:CALCulate:DBM:STATe? – Queries the current dBm state.
:CALCulate:DBM:STATe {ON|OFF|1|0} – Sets the desired dBm state.
Return Format:
The query returns the current dBm state, ON (1) or OFF (0).
Example:
Turn off the dBm operation.
:CALCulate:DBM:STATe OFF
15. :CALCulate:PF?
Command Format:
:CALCulate:PF?
Function:
The command queries the result of current PF operation.