Sense:]resistance:resolution – RIGOL DM3058 User Manual
Page 137

RIGOL
Programming Guide for DM3058/DM3058E
4-44
Parameter:
Name
Type
Range
Default
Refer to Table 4-2
1ppm x range
Explanations:
[SENSe:]CURRent:AC:RESolution {
the desired precision of AC current measurement
[SENSe:]CURRent:AC:RESolution? [MINimum|MAXimum] – Queries the current
precision of AC current measurement.
4-2.
Return Format:
The query returns the current precision of AC current measurement.
22. [SENSe:]RESistance:RESolution
Command Format:
[SENSe:]RESistance:RESolution {
[SENSe:]RESistance:RESolution? [MINimum|MAXimum]
Function:
The commands set and query the precision of 2-wire resistance measurement.
Parameter:
Name
Type
Range
Default
Refer to Table 4-2
3ppm x range
Explanations:
[SENSe:]RESistance:RESolution {
desired precision of 2-wire resistance measurement
[SENSe:]RESistance:RESolution? [MINimum|MAXimum] – Queries the current
precision of 2-wire resistance measurement.
4-2.
Both 2-wire and 4-wire resistance measurements are influenced by this command.
All the commands of 2-wire and 4-wire resistance will have influence over the