Sense:]current:dc:nplc – RIGOL DM3058 User Manual
Page 139

RIGOL
Programming Guide for DM3058/DM3058E
4-46
Function:
The commands set and query the integration time for DC voltage measurement.
The inputted and returned values are multiple of PLC.
Explanations:
[SENSe:]VOLTage:DC:NPLC {0.02|0.2|1|10|100|MINimum|MAXimum} – Sets the
desired integration timeintegration time, MIN = 0.02 PLC, MAX = 100 PLC, and
the default is 1 PLC.
[SENSe:]VOLTage:DC:NPLC? [MINimum|MAXimum] – Queries the current
integration time.
The integration time relates to resolution and measurement rate, refer to Table
4-2.
Return Format:
The query returns the current integration time of DC voltage measurement.
25. [SENSe:]CURRent:DC:NPLC
Command Format:
[SENSe:]CURRent:DC:NPLC {0.02|0.2|1|10|100|MINimum|MAXimum}
[SENSe:]CURRent:DC:NPLC? [MINimum|MAXimum]
Function:
The commands set and query the integration time for DC current measurement.
The inputted and returned values are multiple of PLC.
Explanations:
[SENSe:]CURRent:DC:NPLC {0.02|0.2|1|10|100|MINimum|MAXimum} – Sets
the desired integration time, MIN = 0.02 PLC, MAX = 100 PLC, and the default is
1 PLC.
[SENSe:]CURRent:DC:NPLC? [MINimum|MAXimum] – Queries the current
integration time of DC current measurement.
The integration time relates to resolution and measurement rate, refer to Table
4-2.
Return Format:
The query returns the current integration time of DC current measurement.