Introduction, Ion chamber care, Cleaning – Fluke Biomedical TNT 12000 User Manual
Page 93: Firmware upgrade, Detector

Maintenance
Introduction
4
4-3
Introduction
The TNT 12000 is a calibrated instrument. Fluke Biomedical recommends a calibration
period of one year. To help make sure the TNT 12000 stay in calibration, prevent
mechanical abuse during operation and storage.
W Caution
To prevent damage to the TNT 12000, always operate the TNT
12000 in the specified operation temperature range and
humidity in the specifications.
Ion Chamber Care
When the ion chamber is not connected to a cable, put the dust cap on the connector.
The dust cap helps keep contaminates out of the connector, decrease damage to the
connector, and assures a long service life without electrical leakage.
Install the dust cap on the Dosimeter Detector input connector when it is not connected to
an ion chamber.
Cleaning
Clean the TNT 12000 external surfaces with a damp cloth and weak detergent.
W Caution
To prevent damage to the TNT 12000 or an unwanted effect on
performance, do not apply liquids as a spray or put in liquids.
The unit is not waterproof.
Firmware Upgrade
The Display and Detectors contain firmware that is upgradeable. New firmware is
available through the Fluke Biomedical web site. After new firmware is downloaded
from the web site to your PC, the Display or Detector is connected to the PC to receive
the new firmware.
Note
The Excel Add-In and TNT 12000 Vendor Class Driver must be installed to
do a firmware upgrade.
Go to www.FlukeBiomedical.com and download the upgrade file to your PC.
Detector
1. Double-click the downloaded executable file for the Detector.
2. When the Device Validation window shows, connect the Detector to the PC
through the Mini-B to Type-A USB cable.
3. Push O on the Detector.
4. When the upgrade button on the Device Validation window is activated, click the
Upgrade button.
5. While the firmware downloads to the Detector, a progress bar in a Device Upgrade
window shows in the PC display. LEDs on the Detector blink while data moves to
the Detector.
Note
Do not click on the close button in the upper-right hand corner of the
Device Upgrade window during the download. This will abort the
download.