Dosimeter status – Fluke Biomedical TNT 12000 User Manual
Page 65

Operation
How to Use the Dosimeter Detector (TNT 12000 DoseMate)
2
2-37
fct32.bmp
3. Use A, B, C, and D to highlight a number button and push E to add the
number to the entry. Use a number between 1 and 10 for an offset.
Note
The unit of measurement for Temperature is
°
C and pressure is
mmHg
. This
is different than the units set in the Unit Setup screen.
4. Use A or B to highlight the Save button and push E. The offset value is
automatically summed with the measured value and the user set button shows the
new value.
Use the same steps to set an offset pressure value.
To set a user set temperature value:
1. Use B or A to highlight the button to the right of the Internal Temperature label
and under the User Set label.
2. Push E to open the NUMERIC screen.
3. Use A, B, C, and D to highlight a number button and push E to add the
number to the entry. Use a number in the range between the measured value – 10 and
the measured value + 10.
4. Use A or B to highlight the Save button and push E. The user set value is
automatically summed with the measured value and the user set button shows the
new value.
Dosimeter Status
To examine the status of the Dosimeter Detector:
1. Use A or B to highlight the Setup button in the MEASUREMENT screen and push
E
.
2. Use C or D to highlight the Status button and push E. The STATUS screen
shows in the Display.