H. ptb information, Introduction, Technical data of type examination by ptb – Fluke Biomedical TNT 12000 User Manual
Page 133: Display and detector units, Tnt 12000wd, Ptb information, Appendix h

H-1
Appendix H
PTB Information
Introduction
The Fluke Biomedical TNT 12000 measurement system, with the TNT 12000WD
detector and the TNT 12000 DoseMate (with the 96035B and 96020C ionization
chambers) is approved for calibration by the Physikalisch-Technischen Bundesanstalt
(PTB) in Germany. The approval holder company is:
Fluke Biomedical
Division of Fluke Electronics Corporation
6920 Seaway Blvd.
Everett, WA 98203
USA
The Approval sign is:
23.04
11.03
Technical Data of Type Examination by PTB
Display and Detector Units
Dimensions ........................................................... 5.24 cm x 11.43 cm x 4.45 cm
Weight
TNT 12000D Display .......................................... 347 g
TNT 12000WD X-Ray Detector .......................... 560 g
TNT 12000 DoseMate ........................................ 392 g
TNT 12000WD
The following measurement and calibration nominal ranges apply.
Ranges
Dose ................................................................... 12
μ
Gy – 999 Gy (R/F & MAM)
Dose Rate........................................................... 70
μ
Gy/s – 86 mGy/s