How to change profile parameters – Fluke Biomedical TNT 12000 User Manual
Page 40

TNT 12000
Users Manual
2-12
Table 2-3. Default Parameter Values for Each Measurement Mode
Measurement Mode
Parameter
Radio Mammo Fluoro Dental
Dose Units
R (Roentgen)
R (Roentgen)
R (Roentgen)
R (Roentgen)
Time Units
msec (milliseconds) msec (milliseconds)
NA
msec (milliseconds)
Dose Rate Units
NA
NA
R/min
NA
% kV
90 %
90 %
NA
0 %
Fluoro Type
NA
NA
CONT (continuous)
NA
Target/Filter W/AI Mo/Mo W/AI W/AI
Delay
0 msec
0 msec
NA
0 msec
How to Change Profile Parameters
The four manual profiles, four auto profiles, and all user-defined profiles contain
parameters that can be changed. After parameters are changed and then used to set the
Detector, those changes redefine the profile in memory.
To change the parameters in a profile:
1. Highlight the Settings button in the MODE SETUP screen or the Edit Profile
button in the MEASUREMENT screen.
2. Push E.
The SETTING screen shows in the Display.
fct218.png
Note
The manual settings screen for the Radio mode is shown in the above
example. The banner text shows the mode selection (RADIO, MAMMO,
FLUORO, or DENTAL) and the profile (AUTO, MANUAL, DEFAULT or
user-defined). See Table 2-2 for the parameters that are available for each
measurement mode.
3. Use C or D to highlight the parameter you want to change and push E to open
the drop-down list.
4. Use C or D to highlight a value in the list and push E to set the parameter to