Fluke Biomedical TNT 12000 User Manual
Page 50

TNT 12000
Users Manual
2-22
How to Place the Ion Chamber for Horizontal X-ray Beams
To setup the ion chamber on the test stand, see Figure 2-11 and do the steps that follow.
X-Ray Beam
Cable Clip
Filter Tray
Rod Clamp
HVL Filter
Baseplate
Ion Chamber
Ion Chamber
Bracket
Rod Clamp
fct21.eps
Figure 2-11. Test Stand Configuration for Horizontal Tube
1. Put the short square rod into the baseplate and tighten the thumbscrew.
2. Put the rod clamp on the short square rod and move it to approximately the middle of
the rod. Tighten the rod clamp thumbscrew to the short rod.
3. Put the long square rod into the remaining rod clamp hole and tighten the rod clamp
thumbscrew. This makes a horizontal square rod.
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)