How to setup a wireless connection, How to setup a usb connection, Multiple detector connections – Fluke Biomedical TNT 12000 User Manual
Page 32: Communication settings

TNT 12000
Users Manual
2-4
How to Setup a Wireless Connection
To connect the Display to the Detector through a wireless connection:
1. Push O on the Detector.
2. Push O on the Display.
Go to the Communications Settings section to complete the communication setup.
How to Setup a USB Connection
To connect the display to a Detector through a USB connection:
1. Connect the ends of the Mini-A to Mini-B USB cable to the USB port of the Display
and the Detector as shown in Figure 2-1.
fct04.eps
Figure 2-1. Wired (USB) Connection
2. Push O on the Detector.
3. Push O on the Display.
Go to the Communications Settings section to complete the communication setup.
Multiple Detector Connections
A maximum of two Detectors can operate with the Display at one time. The X-ray or
Dosimeter Detector can operate with the mAs Detector at the same time. The X-ray
Detector and Dosimeter Detector can not operate with the Display at the same time.
You can make the connections through a combination of USB and/or wireless
connections. With the Display and Detectors turned on, set the combination pair through
the CONNECTION screen in the Communication Settings section.
Communication Settings
When power is applied to the Display, the system does a self-test while a message shows
in the Display. After the self-test, the software and hardware versions show in the
Display.
If the Display senses no Detectors, a No Detector Found screen shows in the Display.
If it senses only one Detector, the communication settings are complete. For the
subsequent step, go to the manual section for the sensed Detector. If it senses more than
one Detector (within rf range and/or connected by wire), the connection screen (see
Figure 2-2) shows in the Display.