Fluke Biomedical TNT 12000 User Manual
Page 120

TNT 12000
Users Manual
D-4
Table D-1. Specifications for IEC 61237 Defined Beam Qualities
kVp
Beam
Quality
First HVL
(mm Al)
Beam
Quality
First HVL
(mm Al)
Beam
Quality
First HVL
(mm Al)
40
RQR 2
1.42
RQA 2
2.2
50
RQR 3
1.78
RQA 3
3.8
60
RQR 4
2.19
RQA 4
5.4
70
RQR 5
2.58
RQA 5
6.8
80
RQR 6
3.01
RQA 6
8.2
90
RQR 7
3.48
RQA 7
9.2
100 RQR
8
3.97
RQA
8 10.1 RQT
8 6.9
120 RQR
9
5.00
RQA
9 11.6 RQT
9 8.4
150 RQR
10
6.57
RQA
10 13.3 RQT
10 10.1
500 – 200 Measurement ranges
Values for ion chambers are calculated using nominal sensitivities: 3.2 x 10
8
R/C.
Table D-2. 500-200 Measurement Ranges
Units Effective
Range
[1]
Threshold
Resolution
Step Size
R 10
μ
- 9999
250
μ
0.1
μ
R/s 10
μ
- 100
250
μ
0.1
μ
R/min
1 m – 6 k
15 m
0.01 m
R/hr
0.1 – 360 k
0.9
0.001
R/pulse
[2]
1 m – 1.67
4.17
μ
0.01
μ
Gy 0.1
μ
- 876 m
2.19
μ
0.001
μ
Gy/s 0.1
μ
- 876 m
2.19
μ
0.001
μ
Gy/min
0.01 m - 52.5
131
μ
0.1
μ
Gy/hr
1 – 3.15 k
7.88 m
0.01m
Gy/pulse
[2]
0.01
μ
- 14.6 m
36.5 n
0.1n
[1]
IEC 61674 effective range at 1 % resolution steps.
[2]
At 60 frames/second (1 to 120 frames/selectable).
Note
Introduction of material, other than free air, behind the Ion Chamber will
cause its response to change due to backscatter.