Fluke Biomedical TNT 12000 User Manual
Page 64

TNT 12000
Users Manual
2-36
How to Adjust the Internal Temperature/Pressure Sensor
The Dosimeter Detector lets you override the internal measured values through an offset
value, or a specified temperature or pressure entry. Before you override the temperature
and pressure values, let the Detector equalize for a minimum of one hour. Use only
calibrated thermometers and barometers to set the temperature and pressure.
To override the temperature and/or pressure values:
Use C or D to highlight the Setup button in the Dosimeter MEASUREMENT screen
and push E. The SETUP screen shows in the Display.
fct254.png
Use C or D to highlight the T & P Sensor Adj button and push E. The TEMP &
PRES SENSOR ADJ screen shows in the Display.
fct253.png
You can override the temperature and pressure values with an offset value or a user set
value. To set an offset temperature value:
1. Use A or B to highlight the button to the right of the Internal Temperature label
and under the Offset label.
2. Push E to open the NUMERIC screen.