How to set units of measurement – Fluke Biomedical TNT 12000 User Manual
Page 60

TNT 12000
Users Manual
2-32
fct250.png
2. Use C or D to highlight the Mode dropdown box and push E.
3. Use C or D to highlight one of the three modes and push E.
4. Use C or D to highlight the Continue button and push E. The mode is set
and the MEASUREMENT screen shows in the Display.
To set Automatic Reset:
1. Use A, B, C, and D to highlight the checkbox next to the Automatic Reset label.
2. Push E to toggle the checkbox.
In automatic reset, the last calculated value is held in the Display until a subsequent
above-threshold exposure occurs, a function changes, or a power down. When exposures
occur at less than one-second intervals, the Detector measures them as one exposure.
How to Set Units of Measurement
The Dosimeter measures radiation in
Roentgens (R)
and Grays (Gy). Current
measurement units are coulombs and amperes. To set the units of measure:
1. Use B or A to highlight the Units button in the Dosimeter MEASUREMENT screen
and push E. The UNITS SETUP screen shows in the Display.
fct251.png
2. Use C or D to highlight the Dose dropdown box and push E.