Fluke Biomedical TNT 12000 User Manual
Page 27

Introduction
Specifications
1
1-15
150 cc Ion Chamber
Units Effective
Range
[1]
Threshold
Resolution Step Size
R
10
μ
to 2
6
μ
0.1
μ
R/s
10
μ
to 2
6
μ
0.1
μ
R/min
0.5 m to 120
0.36 m
5
μ
R/hr
10 m to 7.2 k
21.6 m
0.1 m
R/frame
[2]
0.2
μ
to 33 m
0.1
μ
0.002
μ
Gy
0.1
μ
to 0.02
0.052
μ
0.001
μ
Gy/s
0.1
μ
to 0.02
0.052
μ
0.001
μ
Gy/min
5
μ
to 1.2
3.15
μ
0.05
μ
Gy/hr
0.1 m to 72
0.189 m
0.001 m
Gy/frame
[2]
0.002
μ
to 0.33 m
0.8 n
0.02 n
[1]
IEC 61674 effective range at 1 % resolution steps.
[2]
At 60 frames/s (1 to 120 frames/selectable).
150 cc Low Rate Ion Chamber
Units Effective
Range
[1]
Threshold
Resolution Step Size
R/s
2
μ
to 2
[3]
NA
0.1
μ
R/min
0.1 m to 120
[3]
NA
5
μ
R/hr
2 m to 7.2 k
[3]
NA
0.1
m
R/frame
[2]
0.04
μ
to 33 m
[3]
NA
0.002
μ
Gy/s
0.02
μ
to 0.02
[3]
NA
0.001
μ
Gy/min
1
μ
to 1.2
[3]
NA
0.05
μ
Gy/hr
0.02 m to 72
[3]
NA 0.001
m
Gy/frame
[2]
0.4 n to 0.33 m
[3]
NA
0.02
n
[1]
IEC 61674 effective range at 1 % resolution steps.
[2]
At 60 frames/s (1 to 120 frames/selectable).
[3]
Low Rate effective range at 5 % resolution steps.
Electrical Units
Units Effective
Range
[1]
Threshold
Resolution Step Size
C
1 p to 100 n
0.5 pC
0.01 p
A
1 p to 100 n
250 fA
0.01 p
[1]
IEC 61674 effective range at 1 % resolution steps.