Users Manual
2-28
4. Use C, D, B, and A to highlight a character and push E to construct a name
that identifies the ion chamber definition. You can have a maximum of 15 characters
in the name.
5. Use C or D to highlight the Save button and push E. The ion chamber
definition name is put in memory and the Display shows the ION CHAMBER
SETUP screen.
6. Use C or D to highlight the >> button next to the Cal Factor Name label and
push E. The ALPHANUMERIC SCREEN shows in the Display.
7. Use C, D, B, and A to highlight a character and push E to construct a name
that identifies the cal factor. You can have a maximum of 15 characters in the name.
8. Use C or D to highlight the Save button and push E. The cal factor name is
put in memory and the Display shows the ION CHAMBER SETUP screen.
9. Use C or D to highlight the >> button next to the Cal Factor Value label and
push E. The NUMERIC screen shows in the Display.
fct32.bmp
10. Use A, B, C, and D to highlight a number button and push E to add the
number to the entry. You can have a maximum of seven digits in the range of -
9.999999 to +9.999999.
11. Use B or A to highlight the Save button and push E. The cal factor value is put
in memory and the Display shows the ION CHAMBER SETUP screen.
12. Use B or A to highlight << button next to the e label to enter an exponent for the cal
factor.
13. Use A, B, C, and D to highlight a number button and push E to add the
number to the entry. You can have a maximum of two digits in the range of -20 to
+20 for an exponent.
14. Use B or A to highlight the Save button and push E. The cal factor value is put
in memory and the Display shows the ION CHAMBER SETUP screen.
15. Use B or A to highlight the dropdown list and push E.
16. Use C or D to highlight one of the cal factor units and push E. Table 2-4
shows the available cal factor units.