Fluke Biomedical TNT 12000 User Manual
Page 53

Operation
How to Use the Dosimeter Detector (TNT 12000 DoseMate)
2
2-25
fct241.png
Figure 2-14. Ion Chamber Setup Screen
If No Ion Chamber is highlighted in the ION CHAMBER SETUP screen, use D to
highlight New and push E. Go to the How to Add an Ion Chamber Definition or
Cal Factor section to add an ion chamber.
If an ion chamber definition is in the Detector:
3. Use C and D to highlight the desired ion chamber in the left frame of the screen.
4. Use B to move the highlight to the right frame of the screen.
If there are no Cal Factors for the Ion Chamber, then highlight New and push E.
See the How to Add an Ion Chamber Definition or Cal Factor section to make a cal
factor.
If there is a Cal factor, then:
5. Use C and D to highlight the desired calibration factor and push E.
6. Use B or A to highlight the Continue button and push E.
If Air Density Correction Factor (ADCF) is set to off or if it is on and the
Temperature/Pressure is configured for Internal Temperature/Pressure Compensation
then the Dosimeter Measurements screen shows on the Display. Go to the Dosimeter
Measurements section to continue the measurement procedure.