Fluke Biomedical TNT 12000 User Manual
Page 57

Operation
How to Use the Dosimeter Detector (TNT 12000 DoseMate)
2
2-29
Table 2-4. Cal Factor Units
Cal Factor Units
Description
R/C
Roentgens per Coulomb
R/nC
Roentgens per nano Coulomb
Gy/C
Grays per Coulomb
Gy/nC
Grays per nano Coulomb
Note
This unit definition is independent of the units set for Dose or Dose rate.
17. Use B or A to highlight the Continue button and push E. The Display shows
the ION CHAMBER SETUP screen.
How to Edit an Ion Chamber Definition or Cal Factor
To edit an ion chamber definition:
1. Use B or A to highlight the IC button in the Dosimeter MEASUREMENT screen and
push E. The ION CHAMBER SETUP screen shows in the Display.
2. Use C or D to highlight one of the ion chamber definition names and push E.
3. Use C or D to highlight the Edit button and push E. The EDIT ION
CHAMBER screen shows in the Display.
fct248.png
4. Use A, B, C, and D to highlight the button next to the value you want to edit and
push E. In the case of the dropdown list, highlight the dropdown box and push
E
.
See the How to Add an Ion Chamber Definition or Cal Factor section for the steps on
how to set each value.
5. When the values are set, use C or D to highlight the Cancel button and push
E
.