96035b ion chamber, Tnt 12000 dosemate with 96035b ion chamber, Values of the correction factor k – Fluke Biomedical TNT 12000 User Manual
Page 136

TNT 12000
Users Manual
H-4
Values of the correction factor k
Q
, for the radiation qualities as determined by
measurement at the PTB.
RQR – Series
Beam Quality
IEC 61267
Mean Energy, Air Kerma
(keV)
k
Q
(Q
n
/Q
70
)
RQR 2
26.2
1.019
RQR 3
29.0
1.012
RQR 4
32.0
1.005
RQR 5
34.8
1
RQR 6
37.8
0.996
RQR 7
41.0
0.992
RQR 8
44.2
0.989
RQR 9
50.8
0.984
RQR 10
61.2
0.979
RQA – Series
Beam Quality
IEC 61267
Mean Energy, Air Kerma
(keV)
k
Q
(Q
n
/Q
70
)
RQA 3
37.7
1.021
RQA 4
44.6
1.009
RQA 5
51.0
1
RQA 6
57.4
1.001
RQA 7
62.9
0.997
RQA 8
68.0
0.992
RQA 9
78.4
0.994
RQA 10
93.2
0.996
96035B Ion Chamber
Dimensions (Diameter x Height) ........................... 63.5 mm x 13 mm
TNT 12000 DoseMate with 96035B Ion Chamber
The following measurement and calibration nominal ranges apply.
Ranges
Dose ................................................................... 1
μ
Gy – 180 mGy
Dose Rate........................................................... 1
μ
Gy/s – 180 mGy/s
Nominal ranges of use for other factors
Conventional Diagnostic
Anode
Quality
Range of Tube Voltage
Reference
W
RQR2 - RQR10
40 - 150 kV
IEC 61267
W
RQA3 - RQA10
50 - 150 kV
IEC 61267
Mammography
Anode
Total Filtration
Range of Tube Voltage
PTB Reference
Mo
30
μ
m Mo
20 - 50 kV
MMV 20 - MMV 50
Mo
30
μ
m Mo + 2 mm Al
20 - 50 kV
MMV 20 - MMV 50
Nominal ranges of use for other factors
Factor Nominal
Range
Direction of beam incidence
Within a cone with a half angle of 5
°
to the preferred direction
Ambient temperature
15
°
C to 35
°
C
Ambient pressure
800 hPa to 1060 hPa
Relative humidity
20 % to 75 %, <20 g/m