How to charge the battery – Fluke Biomedical TNT 12000 User Manual
Page 21

Introduction
How to Charge the Battery
1
1-9
How to Charge the Battery
Charging the internal batteries of the detectors and display from mains power can be done
in three different configurations as shown in Figure 1-1.
WCaution
To prevent damage to the TNT 12000, do not leave batteries
unused for an extended period of time. When a battery has not
been used for six months, check the charge status and charge
if necessary.
OR
Power Source
Computer to Power Source
Detector to Display to Power Source
OR
fct10.eps
Figure 1-1. Charging Methods
To charge a Detector or the Display directly from mains power, connect the ac-power
adapter directly to the ac adapter connector on a Detector or the Display. To charge a
Detector and the Display simultaneously, connect the ac adapter to the ac adapter
connection on the Display and then, using the Mini-B to Type-A USB cable, connect the
Detector to the Display.
Note
See the markings on the ends of the cable to identify the Mini-A and Mini-B
ends.
A Detector or Display battery can also be charged through a USB connector on a PC.
Connect the Display or Detector to a PC USB port using the Mini-A to Type-A USB
cable.
Both Detectors and the Display have a battery-status indicator to indicate the condition of
the internal battery. The indicator is a tri-color LED that shows the level of charge on the
internal battery. Table 1-5 explains the battery-status indicator levels.