Fluke Biomedical TNT 12000 User Manual
Page 58

TNT 12000
Users Manual
2-30
How to Delete an Ion Chamber Definition or Cal Factor
To delete an ion chamber definition:
1. Use B or A to highlight the IC button in the Dosimeter MEASUREMENT screen and
push E. The ION CHAMBER SETUP screen shows in the Display.
2. Use C or D to highlight one of the ion chamber definition names and push E.
3. Use C or D to highlight the Delete button and push E. The DELETE IC/CF
screen shows in the Display.
fct249.png
4. Use B or A to highlight the Delete IC button and push E.
Use B or A to highlight the Cancel button and push E if you want to cancel the
delete operation.
To delete a cal factor:
1. Use B or A to highlight the IC button in the Dosimeter MEASUREMENT screen and
push E. The ION CHAMBER SETUP screen shows in the Display.
2. Use C or D to highlight one of the ion chamber definition names.
3. Use B to move to the cal factor list.
4. Use C or D to highlight one of the ion chamber cal factor names and push E.
The DELETE IC/CF screen shows in the Display.
5. Use B or A to highlight the Delete CF button and push E.
Note
If the deleted cal factor is the only cal factor for an ion chamber, then the
ion chamber is also deleted.