Electrical connection data KB version cUL for the USA and Canada
Unit size

240-UL

400-UL

720-UL

Electrical data KB-UL
Nominal voltage (+/-10%)

V

100-120

1N~

100-120

1N~

200-240

2~

Power frequency

Hz

60

60

60

Power plug

NEMA

5-20P

5-20P

6-20P

Nominal power

kW

1.70

1.40

2.30

Overvoltage category acc. to IEC 61010-1

II

II

II

II

Pollution degree acc. to IEC 61010-1

2

2

2

2

Unit fuse 6,3 x 32 mm 250V / super-time-lag TT

Amp

16

external

16

external

--

Over-current release category B, 2 poles

Amp

--

--

16

Additional temperature fuse

class 1 (DIN 12880) internal

1)

Lower values are valid up to an ambient temperature of max. 25 °C / 77°F.

2)

to 98% of the set-point

3)

Use these values for sizing air condition systems

All technical data is specified for unloaded units with standard equipment at an ambient temperature of

+25 °C / 77°F and a power supply voltage fluctuation of +/-10%. The temperature data is determined in

accordance to BINDER factory standard following DIN 12880, observing the recommended wall

clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100%

fan speed.
All indications are average values, typical for units produced in series. We reserve the right to

change technical specifications at any time.

Refrigerating performance decreases while operating the chamber at temperatures < 0 °C /

32°F due to icing of the evaporators. For this reason defrost the chamber regularly, e.g. once a

week.

If the chamber is fully loaded, the specified heating up and cooling down times may vary

according to the load.