Options, 2 ethernet interface (option for kb 240, 400, 720), 3 data logger kit (option) – BINDER KB 23 User Manual
Page 71
KB (E3.1 + E5.1) 11/2014
page 71/100
16. Options
16.1 Communication software APT-COM™ 3 DataControlSystem (option)
The unit is regularly equipped with a serial interface RS 422 (13) that can connect the BINDER
communication software APT-COM™ 3 DataControlSystem. The connection to a computer is established
using the KB interface via an interface converter.
Confirm that the interface mode is correctly set to “Modbus” in the user level (chap. 10).
The actual temperature, and fan speed values are given in adjustable intervals. Programming can be
performed graphically via PC. Up to 30 chambers with RS 422 interface can be cross linked. For further
information, refer to the operating manual of the BINDER communication software APT-COM™ 3.
Pin allocation of the RS 422 interface (13) at the unit rear:
Pin 2:
RxD (+)
Pin 3:
TxD (+)
Pin 4:
RxD (-)
Pin 5:
TxD (-)
Pin 7:
Ground
16.2 Ethernet interface (option for KB 240, 400, 720)
With this option, the chamber is equipped with an Ethernet interface (10) instead of the RS422 interface,
that can connect the BINDER communication software APT-COM™ 3 DataControlSystem. The MAC
Address is indicated below the Ethernet interface. The actual temperature, and fan speed values are
given in adjustable intervals. Programming can be performed graphically via PC. For further information,
refer to the operating manual of the BINDER communication software APT-COM™ 3.
16.3 Data logger kit (option)
The BINDER Data Logger Kit offers an independent long-term measuring system for temperature,
available for different temperature ranges.
The BINDER Data Logger is equipped with a keyboard and a large LCD display, alarm functions and a
real-time function. Measurement data are recorded in the Data Logger and can be read out after the
measurement via the RS232 interface of the Data Logger. It offers a programmable measuring interval
and permits storing up to 64000 measuring values. Reading out is done with the Data Logger evaluation
software. You can give out a combined alarm and status protocol directly to a serial printer.
Data Logger Kit T 220: Measuring sensor for the temperature values of the chamber: Temperature range
-90 °C / 194 °F up to +220 °C / 428 °F.
For detailed information on installation and operation of the BINDER Data Logger, please
refer to the mounting instructions Art. No. 7001-0204 and to the original user manual of the
manufacturer, supplied with the data logger.