page 72/100
16.4 Analog output for temperature (option)
With this option the chamber is equipped with an analog output 4-20 mA for temperature. This output
allows transmitting data to external data registration systems or devices.
The connection is realized as a DIN socket.
With KB (E3.1) size 53 and 115, the DIN socket is located on the unit rear:
With KB (E5.1) size 240, 400 and 720, the DIN socket (7) is located in the right lateral control panel.
ANALOG OUTPUT 4-20 mA DC
PIN 1: Temperature –
PIN 2: Temperature +
KB (E3.1): Temperature range: -5 °C / 23 °F to +100 °C / 212 °F
KB (E5.1): Temperature range: -10 °C / 14 °F to +100 °C / 212 °F
Figure 18: Pin configuration of the DIN socket
A suitable DIN plug is enclosed.
16.5 Water protected internal socket (option from size 53 on – not for KB-UL)
The internal socket is splash proof.
IP system of protection 67 230 V 1N ~ 50-60 Hz
Charge max. 500 W
Maximum permitted operating temperature with this option: 90 °C / 194 °F.
WARNING
Exceeding the permitted maximum temperature.
Electrical hazard.
Danger of death.
Damage to the internal socket.
∅
Do NOT exceed the temperature set-point of 90 °C / 194 °F.
Set the safety controller to 90 °C / 194 °F.
With optional temperature safety device, class 3.3, set the mechanical thermostat
class 3.1 to 90 °C / 194 °F.
Heat emission of electrical devices connected inside the chamber may modify the temperature
range.