Caution, Warning – BINDER KB 23 User Manual
Page 81
![background image](https://www.manualsdir.com/files/767076/content/doc081.png)
KB (E3.1 + E5.1) 11/2014
page 81/100
The cooling incubator KB bears the symbol for the marking of electrical and electronic
equipment manufactured / placed on the market in the EC after 13 August 2005 and be
disposed of in separate collection according to the directive 2002/96/EC on waste electrical
and electronic equipment (WEEE). WEEE marking: crossed-out wheeled bin with solid bar
under.
At the end of the device’s service life, notify the distributor who sold you the device, who will
take back and dispose of the unit according to the directive 2002/96/EC of 27 January 2003
on waste electrical and electronic equipment (WEEE).
CAUTION
Violation against existing law.
∅
Do NOT dispose of BINDER devices at public collecting points.
Have the device disposed of professionally at a recycling company that is certified
according to conversion of the directive 2002/96/EC into national law.
or
Instruct the distributor who sold you the device to dispose of it. The agreements apply
that were agreed with the distributor when purchasing the unit (e.g. his general terms of
payment and delivery).
If your distributor is not able to take back and dispose of the unit, please contact
BINDER service.
Certified companies disassemble waste (used) BINDER equipment in primary substances for recycling
according to directive 2002/96/EC. The devices must be free from toxic, infectious or radioactive
substances in order to eliminate any health hazards to the employees of the recycling companies.
Prior to handing the unit over to a recycling company, it is the user’s responsibility that it is free
from toxic, infectious or radioactive substances.
Prior to disposal, clean all introduced or residual toxic substances from the unit.
Prior to disposal, disinfect the unit from all sources of infection. Be aware that sources of
infection may also be located outside the inner chamber.
If you cannot safely remove all sources of infection and toxic substances from the unit, dispose
of it as special waste according to national law.
Fill out the contamination clearance certificate (chap. 21) and enclose it with the unit.
WARNING
Contamination of the device with toxic, infectious or radioactive substances.
Danger of intoxication.
Danger of infection.
∅
NEVER take a unit contaminated with toxic substances or sources of infection for
recycling according to directive 2002/96/EC.
Prior to disposal, remove all toxic substances and sources of infection from the unit.
A unit from which all toxic substances or sources of infection cannot be safely removed
must be considered as “special” waste according to national law. Dispose of it
accordingly.
The refrigerant used R134A (1,1,1,2-tetrafluorethane) is not inflammable at ambient pressure. It must not
be exposed to the environment. In Europe, recovery of the refrigerant R134A (GWP 1300) is mandatory
according to regulation No. 842/2006/EC. Ensure the compliance with the applicable legal requirements
regarding qualification of staff, disposal, and documentation.