10. User level
In this menu the following parameters can be se (in brackets the corresponding abbreviated information
given in display 2):
•
Unit address (Address)
Setting of controller address (1 to 254) for operation with the communication software APT-COM™.
•
User code (User-cod)
Modification of the user code setting (factory setting “0001”) for access to the user level and the
program editor.
Keep in mind any modification of the user code. Without a valid user code it is no longer
possible to access these levels.
•
Safety controller mode (Saf.mode)
The set-point mode can be entered as ...
Limit: Maximally permitted temperature as absolute value (example: temperature set-point 37 °C /
98.6°F, limit setting to 39 °C / 102.2°F)
Offset: Maximally permitted temperature exceeding of the actual set-point temperature (e.g. 2 °C).
This maximally permitted temperature automatically changes with every change of the temperature
set-point.
•
Safety controller set-point (Saf.setp)
The temperature set-point setting of the safety controller (over temperature safety device class 3.1) is
shown and can be changed. An absolute value (e.g. 40 °C / 104°F) in case of setting “limit” or a
relative value (e.g. 2 °C) in case of setting “Offset” can be entered.
Regularly check the setting of the set-point mode and of the over temperature safety
device set-point entered in this mode for the temperature value in Fixed value operation or
for the maximum temperature value of the selected temperature program in Program
operation mode.
•
Decimal point position (Decimal)
Selection if integer values or one position after the decimal point can be entered. The integer
representation is shown in Display 2 (set-point entry) while the actual value in Display 1 is always
displayed with one decimal point.
•
Audio Alert (Buzzer)
Inactive: no audible signal (buzzer) in case of an alarm event.
Active: in case of an alarm event (see chap. 13.2) an audible signal (buzzer) will sound. It can be reset
by pressing the “
EXIT” button.
•
Selection of controller menu language (Language)
German, English, or French can be selected.
•
Counter of operating hours (Oper.hs)
Information about the number of operating hours currently reached or since the last reset. (no setting,
display only).
•
Max. number of operating hours (Op.limit)
Entry of a limit number of operating hours, i.e., the maximum number of operating hours that can be
run. Maximum setting: 9999. Reaching the limit has no effect.
•
Reset operating hours (Op.back)
Reset operating hours to zero.