6 intended use, Danger – BINDER KB 23 User Manual
Page 14
![background image](https://www.manualsdir.com/files/767076/content/doc014.png)
KB (E3.1 + E5.1) 11/2014
page 14/100
1.6 Intended use
Cooling incubators KB are suitable for exact conditioning of harmless materials. Because of their precise
temperature accuracy these devices are especially useful for cultivation of microorganisms with a narrow
temperature optimum in a range of 4 °C / 39.2°F to 37 °C / 98.6°F. Main fields of application are tests of
long-term storage (e.g. at 4 °C / 39.2°F), refrigerated incubation between 20 °C / 68°F and 25 °C / 77°F
and incubation at 37 °C / 98.6°F (also with additional introduction of heat) or with alternating temperatures
(e.g. 37 °C / 98.6°F and 4 °C / 39.2°F).
A mixture of any component of the charging material with air must NOT be explosive. The operating
temperature must lie below the flash point or below the sublimation point of the charging material. Any
component of the charging material must NOT be able to release toxic gases.
Other applications are not approved.
Cooling incubators KB are not classified as medical devices as defined by the Medical Device
Directive 93/42/EEC.
Following the instructions in this operating manual and conducting regular maintenance work
(chap. 17.1) are part of the intended use.
DANGER
Explosion or implosion hazard.
Danger of poisoning.
Danger of death.
∅
Do NOT introduce any substance combustible or explosive at working temperature into
the constant climate chamber.
∅
NO explosive dust or air-solvent mixture in the inner chamber.
∅
Do NOT introduce any substance which could lead to release of toxic gases.
The charging material shall not contain any corrosive ingredients that may damage the
machine components made of stainless steel, aluminum, and copper. Such ingredients
include in particular acids and halides. Any corrosive damage caused by such ingredients is
excluded from liability by BINDER GmbH.
WARNING: If customer should use a BINDER chamber running in non-supervised continuous
operation, we strongly recommend in case of inclusion of irrecoverable specimen or samples
to split such specimen or samples and store them in at least two chambers, if this is feasible.