beautypg.com

2 set-point entry for program operation – BINDER KB 23 User Manual

Page 35

background image

KB (E3.1 + E5.1) 11/2014

page 35/100


Program table corresponding to the diagram (with setting “Step”):

Section


SEC

Temperature

set-point

[ °C]

TEMP

Section length

[hh.mm]

TIME

Fan speed

[%]

FAN

Operation lines

*

O.LINE

S01

40

00:30

50

000

S02

60

01:30

100

000

S03

80

01:00

100

000

S04

20

03:20

100

000

* Only with option zero-voltage relay outputs via operation lines, see chap. 16.4.
The values of such a program table can now be entered to the RD3 program controller (chap. 8.2).

8.1.3 General notes on programming of temperature transitions

If the tolerance limits set in the user level (chap. 10) are exceeded, the program is halted until the actual

temperature value returns to within the tolerance range. During this program interruption, the LED (3d)

flashes. Therefore, the duration of the program might be extended due to the programming of tolerances
Programming is saved even in case of a power failure or after turning off the unit.

KB 23: Cooling power decreases when the fan speed is reduced to 20% or below.

After program rundown the controller returns to Fixed value operation showing Normal Display and

equilibrates to the temperature value previously entered in fixed value entry mode.

Before starting the program, check the set-point value entered in Fixed value entry mode.

After program rundown temperature will equilibrate to this value.

Deactivate the week program timer (factory setting, setting in the user level, chap. 10) before

starting a program.

8.2 Set-point entry for program operation

From Normal Display the program editor is accessed by pressing button X/W for 5 sec. Then enter the

set-points one after the other in all program sections of a selected program.
You can enter two programs with up to 10 sections each or one program with up to 20 sections (setting in

the user level, chap. 10).
In order to avoid incorrect programming the values of the program course should be entered into a table

(template in chap. 8.3).
Example program table (with default setting “Ramp”):

Section

SEC

Temperature set-

point

[ °C]

TEMP

Section length

[hh.mm]

TIME

Fan speed

[%]

FAN

Operation lines

*

O.LINE

S01

40

00:30

50

000

S02

60

01:30

100

000

S03

90

01:00

100

000

S04

90

03:20

100

000

S05

20

00:01

100

000

* Only with option zero-voltage relay outputs via operation lines, see chap. 16.4.

This manual is related to the following products: