3 sending the unit back to binder gmbh, Caution – BINDER KB 23 User Manual
Page 78
KB (E3.1 + E5.1) 11/2014
page 78/100
With every decontamination method, always use adequate personal safety controls.
In case of contamination of the interior by biologically or chemically hazardous material, there are two
possible procedures depending on the type of contamination and charging material.
(1) Spray the inner chamber with an appropriate disinfectant.
Before start-up, the unit must be absolutely dry and ventilated, as explosive gases may form during
the decontamination process.
(2) If necessary, have strongly contaminated inner chamber parts removed by an engineer for cleaning,
or have them exchanged. Sterilize the inner chamber parts in a sterilizer or autoclave.
In case of eye contact, the disinfectant spray may cause eye damage due to chemical
burns. Follow the operating instructions and safety hints labeled on the bottle of the
disinfectant spray.
Recommended precautions: To protect the eyes use sealed protective goggles.
CAUTION
Eye contact.
Eye damage due to chemical burns.
∅
Do NOT empty into drains.
Wear protective goggles.
After using the disinfectant spray, allow the unit to dry thoroughly, and aerate it sufficiently.
17.3 Sending the unit back to BINDER GmbH
If you return a BINDER product to us for repair or any other reason, we will only accept the product upon
presentation of an authorization number that has previously been issued to you. An authorization number
will be issued after receiving your complaint either in writing or by telephone
prior to your sending the
BINDER product back to us. The authorization number will be issued following receipt of the information
below:
•
BINDER product type and serial number
•
Date of purchase
•
Name and address of the dealer from which you bought the BINDER product
•
Exact description of the defect or fault
•
Complete address, contact person and availability of that person
•
Exact location of the BINDER product in your facility
•
A contamination clearance certificate (chap. 21) must be faxed in advance
The authorization number must be applied to the packaging in such a way that it can be easily recognized
or be recorded clearly in the delivery documents.
For security reasons we cannot accept a unit delivery if it does not carry an authorization
number.
Return address:
BINDER GmbH
Abteilung Service
Gänsäcker 16
78502 Tuttlingen
Germany