4.2 Electrical connection
The cooling incubators KB are supplied ready for connection.
•
The cooling incubator KB (E3.1) comes with a fixed power connection cable that has a length of 1800
mm / 5.9 ft and is equipped with a semi time-lag fine-wire fuse.
•
The cooling incubator KB (E5.1) comes with a fixed power connection cable that has a length of 1800
mm / 5.9 ft.
Model
Power plug
Voltage
+ /-10 %
Current
type
Power
frequency
Unit fuse
KB 23 (E3.1)
Shock-proof plug
230 V
1N~
50/60 Hz
10 Amp
KB 53 (E3.1)
KB 115 (E3.1)
Shock-proof plug
230 V
1N~
50 Hz
10 Amp
KB 23-UL (E3.1)
KB 53-UL (E3.1)
KB 115-UL (E3.1)
NEMA 5-15
115 V
1N~
60 Hz
12,5 Amp
KB 240 (E5.1)
Shock-proof plug 200 V to 240 V
1N~
50 Hz
16 Amp
KB 400 (E5.1)
KB 720 (E5.1)
Shock-proof plug 200 V to 240 V
1N~
50/60 Hz
16 Amp
KB 240-UL (E5.1)
NEMA 5-20P
100 V to 120 V
1N~
60 Hz
16 Amp
KB 400-UL (E5.1)
NEMA 5-20P
100 V to 120 V
1N~
50/60 Hz
16 Amp
KB 720-UL (E5.1)
NEMA 6-20P
200 V to 240 V
2 ~
50/60 Hz
16 Amp
•
The socket must also provide a protective conductor.
•
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the unit’s type plate (left unit side, bottom right-hand, see chap. 1.4).
•
When connecting, please observe the regulations specified by the local electricity supply company as
well as the VDE directives (for Germany). We recommend the use of a residual current circuit breaker.
•
Pollution degree (acc. to IEC 61010-1): 2
•
Over-voltage category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 20.4).
To completely separate the unit from the power supply, you must disconnect the power plug.
Install the unit in a way that the power plug is easily accessible and can be easily pulled in case
of danger.