BINDER KB 23 User Manual
Page 37
KB (E3.1 + E5.1) 11/2014
page 37/100
Next step – set-point entry in the desired program sections:
Basic entry principle: Access the parameters of individual program sections with button X/W one after
the other. Enter the values of the individual parameters with the arrow keys. A value flashes once after 2
seconds thus indicating that it has been taken over to the controller. If several parameters are to be
skipped (e.g. in order to change a parameter in a posterior program section), the parameters can be
rapidly jumped over by holding down the X/W key. If no button is pressed for more than 120 sec the
controller toggles back to Normal Display. The program entered to this point remains stored.
Press key
Display 1 shows
e.g. 9.0
C
(actual temperature set-point)
Display 2 shows
alternating
S01: TEMP 9.0
(variable: temperature in °C)
CONTINUE X/W
(information: go on with X/W)
Enter temperature set-point of S01 in °C
using arrow keys
Press key
Display 1 shows
e.g. 00.01
(actual section length set-point)
Display 2 shows
alternating
S01: TIME 00:10
(variable: section length in hh:mm)
CONTINUE X/W
(information: go on with X/W)
Enter section length set-point of S01 in
hh.mm using arrow keys
Value is shown in both displays.
Press key
Display 1 shows
e.g. 50
(actual fan speed set-point)
Display 2 shows
alternating
S01:FAN 50
(variable: fan speed in %)
CONTINUE X/W
(information: go on with X/W)
Enter fan speed set-point of S01 in % using
arrow keys
Value is shown in both displays.
Press key
Only with option zero-voltage relay outputs via operation lines (chap. 16.4):
Display 1 shows
e.g. 000
(actually set switching state)
Display 2 shows
alternating
S01:O.LINE 000
(variable: switching state)
CONTINUE X/W
(Information: go on with X/W)
Enter switching state using arrow keys
Press key