4 kb (e3.1) technical data – BINDER KB 23 User Manual
Page 86
KB (E3.1 + E5.1) 11/2014
page 86/100
20.4 KB (E3.1) technical data
Unit size
23
53
115
Exterior dimensions
Width
mm / inch
433 / 17.05
634 / 24.96
834 / 32.83
Height (incl. feet)
mm / inch
618 / 24.33
837 / 32.95
1022 / 40.24
Depth
mm / inch
516 / 20.31
576 / 22.68
646 / 25.43
incl. door handle, I-panel, connection
mm / inch
73 / 2.87
100 / 3.94
100 / 3.94
Wall clearance
mm / inch
100 / 3.94
100 / 3.94
100 / 3.94
Wall clearance door open
mm / inch
100 / 3.94
160 / 6.30
160 / 6.30
Steam space volume
l / cu.ft.
36 / 1.27
77 / 2.72
158 / 5.58
Interior dimensions
Width
mm / inch
222 / 8.74
400 / 15.75
600 / 23.62
Height
mm / inch
330 / 12.99
400 / 15.75
480 / 18.90
Depth
mm / inch
277 / 10.91
330 / 12.99
400 / 15.75
Interior volume
l / cu.ft.
10 / 0.35
53 / 1.9
115 / 4.1
Number of racks standard / max.
2/3
2/4
2/5
Load per rack
Kg / lbs
12 / 26
15 / 33
20 / 44
Permitted total load
Kg / lbs
25 / 55
40 / 88
50 / 110
Weight (empty)
Kg / lbs
44 / 97
72 / 159
105 / 232
Number of door(s)
number
1
1
1
Inner glass door(s)
number
1
1
1
Temperature data
Temperature range 1)
°C
°F
0 to +100
4 to 212
-5 to +100
23 to 212
-5 to +100
23 to 212
Temperature fluctuation
at 5 °C
± K
0.2
0.1
0.1
at 40 °C
± K
0.2
0.1
0.1
Temperature uniformity
(variation)
at 5 °C
± K
0.7
0.6
0.4
at 25 °C
± K
0.3
0.2
0.1
at 40 °C
± K
0.3
0.3
0.2
Heating up time 2)
to 40 °C
min
11
5
9
Cooling down time from 40 °C to 5 °C
min
61
58
83
Recovery time after doors
were open for 30 sec. 2)
at 5 °C
min
5
4
5
at 40 °C
min
2
1
1
Electrical data KB
IP System of protection acc. to EN 60529
20
20
20
Nominal voltage (±10 %)
V
230 1N~
230 1N~
230 1N~
Power frequency
Hz
50/60
50
50
Power plug
shock proof plug
Nominal power
kW
0.34
0.46
0.46
Energy consumption 3)
at 5 °C
Wh/h
50
260
222
at 40 °C
Wh/h
60
215
115
Overvoltage category acc. to IEC 61010-1
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
Unit fuse 5x20 mm 230V / middle-time-lag
M
Amp
10
10
10
external
external
external