Program start level, Caution – BINDER KB 23 User Manual
Page 41

KB (E3.1 + E5.1) 11/2014
page 41/100
The following section (in our example now S03) is shown:
Display 1 shows
e.g. 03
(actual selection of the section: S03)
Display 2 shows
alternating
P01:S03
(program section can be selected)
CONTINUE X/W
(information: set-point entry with X/W)
Press key EXIT
or wait 120 sec
Controller returns to Normal Display
If you delete a program section which is followed by further sections, those following move up
in place of the deleted section.
In our example, section S03 has been deleted. If sections S04, S05, etc. have been programmed earlier,
they will now replace the preceding sections, i.e., S04 is now called S03 etc.
Deletion leads to overwriting the section by the following one. It is therefore not possible to temporarily
inactivate a program section. To enter a section later to a program, all the sections following the new one
must be entered again.
9.
Program start level
Before starting the program, check the set-point entered in Fixed value operation mode. After end of the
program, the temperature will equilibrate to this value.
CAUTION
Too high or too low temperature after the program ends.
Damage to the charging material.
Check the set-point of Fixed value operation and if necessary adapt it.
After the program ends, the temperature will equilibrate to the set-point entered in Fixed value operation
mode. If the week program timer is active, another set-point (SP2) might be targeted according to
programming. Deactivate the week program timer before starting the program (default setting, setting in
the User level, chap. 10).
CAUTION
Too high or too low temperature after the program ends.
Damage to the charging material.
Deactivate the week program timer before starting the program.
In the first step a program is selected. This is on condition that a program has been entered previously
(chap. 8.2) and that program type “2 programs with 10 sections each” has been selected in the user level