BINDER KB 23 User Manual
Page 87

KB (E3.1 + E5.1) 11/2014
page 87/100
Electrical connection data KB version cUL for the USA and Canada
Unit size
23
53
115
Electrical data KB-UL
Nominal voltage (±10%)
V
115 1N~
115 1N~
115 1N~
Power frequency
Hz
60
60
60
Power plug
NEMA
5-15P
5-15P
5-15P
Nominal power:
kW
0.34
0.36
0.46
Overvoltage category acc. to IEC 61010-1
II
II
II
Pollution degree acc. to IEC 61010-1
2
2
2
Unit fuse 6,3 x 32 mm
250V / super-time-lag TT
Amp
12.5
12.5
12.5
external
external
external
Additional temperature fuses
class 1 (DIN 12880) internal
Legend:
1)
at an ambient temperature of max. 25 °C / 77 °F.
2)
to 98% of the set-point
3)
Use these values for sizing air condition systems
All technical data is specified for unloaded units with standard equipment at an ambient temperature of
+25 °C / 77 °F and a power supply voltage fluctuation of ±10. The temperature data is determined in
accordance to BINDER factory standard following DIN 12880, observing the recommended wall
clearances of 10 % of the height, width and depth of the inner chamber. Technical data refers to 100% fan
speed.
All indications are average values, typical for units produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up and cooling down times may vary
according to the load.