11.3 Proceeding in detail

1. Settings in the user level:

Setting the safety controller to “Limit” and 3 °C above the maximum temperature value of the

program

Activating the week program timer

Checking and, if necessary, setting the real time clock

Normal Display

Display 1 shows

e.g. 19.8

(actual temperature value)

Display 2 shows e.g. 15.05.06 13:52 - -

(actual date and time, actual switching state of week

program timer channel 1: Off, channel 2: Off)

Press down key

for 5 sec

Display 1 shows

e.g. 0000

Display 2 shows PROGRAM EDITOR

(you are in the program editor)

Press down key

for 5 sec

Display 1 shows

0000

Menu visible only if week program timer is activated.

Display 2 shows WEEK PROG. EDITOR

(you are in the week program editor)

Press down key

for 5 sec

Display 1 shows

0000

Display 2 shows

USER – LEVEL

(you are in the user level)

Press program key

Display 1 shows

0000

Display 2 shows UserCod?

0000

(enter user code, display flashes)

Enter the user code using arrow keys

e.g.

0001 (basic setting,

or the valid code in case it has been

previously changed in this menu).

Value is shown in both displays.

Automatically forward after 2 sec

Display 1 shows

1

(actual address)

Display 2 shows

Address

1

(entry of unit address)

(actual address: 1)

Press several times key

until

Saf.mode appears:

Display 1 shows

0000

(no function)

Display 2 shows Saf.mode: Limit.

(select set-point mode of safety controller)

(actual setting: „limit“)

Select “

Limit” using arrow keys

Setting is shown in display 2.

Press key

Display 1 shows

23

(actually set temperature value)

Display 2 shows

Saf.setp

23

(temperature value of over temperature safety device)

(actual value: 23 °C / 73.4°F)

Enter the temperature value of

23 in °C

using arrow keys

Value is shown in both displays.