Using the External Trigger and Analog Outputs

High Speed Power Meter

83

Analog Output (IQS-1610, IQS-1620)

Connecting a TTL Source to the External Trigger

A synchronizing signal from a signal generator or from a control circuit may
be connected to the external trigger input of the your high-speed power
meter module if it does not exceed TTL levels. The illustration below
shows a typical connection setup for a signal generator.

Analog Output (IQS-1610, IQS-1620)

The analog output provides an electrical signal proportional to the optical
power input on the detector (which is the measured value shown in the
display of the Instrument function tab). It is available for channel 1 of both
single- and dual-channel models.

The analog output signal is the redirection of the signal after the first
amplification stage. Do not use the analog output in Autorange mode (see
Setting the Measurement Range
on page 32) be
cause the gain scales of the
first amplification stage are constantly changed to optimize
signal-to-noise ratio.

4.000 000 000n GHz

6.00 dBm

2-Ch Power Meter

IQS-1600

EXT.

TRIG

ANALOG OUT

C1

C2

Signal generator

IQS-1600

Optical power input

Optical input

Optical output

External trigger input

DUT

(e.-o. component)