:SENSe[1..n]:POWer[:DC]:RANGe:AUTO

Description

This command enables or disables the

automatic power measurement range

(Autorange) for the currently selected channel.

Syntax

:SENSe[1..n]:POWer[:DC]:RANGe:AUTO

Parameter(s)

AutoRangeState:
The program data syntax for

is defined as a

element. The special forms

ON and OFF are accepted on input for increased

readability. ON corresponds to 1 and OFF

corresponds to 0.

Set the , where:

0 or OFF -Disables the AutoRange

1 or ON -Enables the AutoRange

Example(s)

SENS:POW:RANG:AUTO 1

Notes

The Autorange function can be performed in

Continuous acquisition mode only. When

Autorange is deactivated, LR1 is activated by

default.

See Also

SENSe:POWer:[DC]:RANGe:AUTO?

SENSe:POWer:[DC]:RANGe:HIGH

SENSe:POWer:[DC]:RANGe:LOWer

SENSe:POWer:[DC]:RANGe:SCALe?

SENSe:POWer:[DC]:RANGe:SCALe:LIST?