Performing Acquisitions

High Speed Power Meter

63

Selecting the Acquisition Mode

Setting Trigger Acquisition Parameters

Conditional data acquisition can be performed using a triggered
acquisition, meaning that data recording begins when a specified
condition is met. Different trigger conditions are available, which are
explained in the following table, where A and/or B represent the channel
on which the condition is to be met, and x and/or y represent the desired
power level threshold.

Trigger

Description

A > x

Acquisition will start when measured power is greater

than the specified x value.

A < x

Acquisition will start when measured power is lower

than the specified x value.

A > x AND A < y Acquisition will start when measured power in

channel A is inside the interval set by the values

x and y.

A or B < x

Acquisition will start when power in either channel A
or channel B is lower than the specified x value.

x < A or B < y
(multichannel

power meters

only)

Acquisition will start when measured power in

channel A goes over a specified value (A > x).
OR
Acquisition will start when measured power in

channel B falls under a specified value (B < y).

A < B
(multichannel

power meters

only)

Acquisition will start when measured power of

channel A is lower than measured power of channel B.

External trigger

Acquisition in single rate will be triggered by an

external signal, whether on positive or negative TTL

voltage transitions (SeeExternal Trigger on page 82

for

additional information on this feature).