SCPI Command Reference

220

IQS-1600

Product-Specific Commands—Description

:TRIGger[1..n][:SEQuence]:SOURce?

Description

This query returns the trigger condition's source.

Syntax

:TRIGger[1..n][:SEQuence]:SOURce?

Parameter(s)

None

Response Syntax

Response(s)

TriggerSource:
The response data syntax for is

defined as a

element.

The current trigger source, where:

EXTernal - is the external trigger

INTernal1 - is channel 1

INTernal2 - is channel 2 (IQS-1620 or IQS-1640)

INTernal3 - is channel 3 (IQS-1640) or virtual

channel 1(IQS-1620)

INTernal4 - is channel 4 (IQS-1640)

INTernal5 - is virtual channel 1 (IQS-1640)

INTernal6 - is virtual channel 2 (IQS-1640)

Example(s)

TRIG:SEQ:SOUR?

See Also

TRIGger[:SEQuence]:SOURce

TRIGger[:SEQuence]:LEVel?

TRIGger:POSition

TRIGger:POSition:CATalog?

TRIGger[:SEQuence]:STATe

TRIGger[:SEQuence]:SLOPe