SCPI Command Reference

192

IQS-1600

Product-Specific Commands—Description

:SENSe[1..n]:POWer[:DC]:UNIT

Description

This command changes the absolute

measurement display unit: dBm or W for the

specified channel.

Syntax

:SENSe[1..n]:POWer[:DC]:UNITDBM|W

Parameter(s)

Unit:
The program data syntax for the first parameter is

defined as a

element. The allowed

DATA> elements for this parameter are: DBM|W.

Power unit in dBm or Watt, where:

DBM - sets the power display to dBm

W | WATT - sets the power display to Watt

(pW,nW,uW,Mw…)

Example(s)

SENS:POW:UNIT W

Notes

This command is not executed if a data

acquisition is in progress. In that case, the

"Acquisition in progress" message will be

returned.

See Also

SENSe:POWer[:DC]:UNIT?

UNIT:POWer

UNIT:POWer?