beautypg.com

Initiate[1..n]:auto – EXFO IQS-1600 High-Speed Power Meter for IQS-500/600 User Manual

Page 139

background image

SCPI Command Reference

High Speed Power Meter

131

Product-Specific Commands—Description

:INITiate[1..n]:AUTO

Description

This command starts or stops an acquisition

using the number of points set with the

TRAC:POIN command, and the sampling rate set

with the SENS:FREQ[:CONT] or

SENS:FREQ:NCON commands. To start an

acquisition in non-continuous mode, no channel

must be set to auto scale.

Syntax

:INITiate[1..n]:AUTO,CONT|

NCONt

Parameter(s)

³

StartStop:

The program data syntax for is

defined as a element.

The special forms ON and OFF are

accepted on input for increased readability. ON

corresponds to 1 and OFF corresponds to 0.

This parameter changes the state of an

acquisition:

0 or OFF - Stops the acquisition.

1 or ON - Starts the acquisition.

³

AcqType:

The program data syntax for the second

parameter is defined as a

PROGRAM DATA> element. The allowed

elements for

this parameter are: CONT|NCONt.