beautypg.com

Measuring insertion loss, Typical applications – EXFO IQS-1600 High-Speed Power Meter for IQS-500/600 User Manual

Page 104

background image

Typical Applications

96

IQS-1600

Measuring Insertion Loss

Measuring Insertion Loss

To accurately measure the insertion loss of a fiber-optic component, use a
light source (such as EXFO’s IQS-2100 Light Source) and your high-speed
power meter.

To measure the insertion loss:
1.
Perform an offset nulling of the detector(s) if needed.

2. Make sure that the source is turned off.

3. Connect the source to the power meter using two appropriate test

jumpers and a bulkhead connector as shown below. This setup will be
used to record a reference value.

4. Adjust the source and power meter to the appropriate wavelength.

5. Activate the source and select a suitable power output.

Light Source

IQS-2100

ACTIVE

Bulkhead connector

IQS-1600 Power Meter

IQS-2100 Light Source

2-Ch Power Meter

IQS-1600

EXT.

TRIG

ANALOG OUT

C1