beautypg.com

12 typical applications – EXFO IQS-1600 High-Speed Power Meter for IQS-500/600 User Manual

Page 101

background image

High Speed Power Meter

93

12 Typical Applications

For accurate, significant, and repeatable power measurements, it is
important to consider the following points:

³

Connectors, fiber ends, ports, and detectors should be clean at all
times.

³

A null measurement should be performed prior to each user session or
whenever there is a significant change in ambient temperature or
environmental conditions. To perform a null measurement, see Nulling
Offsets
on page 38.

³

The optical source should be stable.

³

Appropriate test jumpers, connectors, and adapters must be used.

³

Once a reference has been taken, make sure that the setup remains
stable. To take a reference, see Measuring Relative Power on page 43.

³

Since optical losses vary according to the launch and receive
conditions and fiber type, controlling these conditions with mode
filters and strippers is essential.