SCPI Command Reference

High Speed Power Meter

207

Product-Specific Commands—Description

:TRACe[1..n]:POINts

Description

This command sets the number of points to store

for all traces. The maximum number of points

that can be acquired depends on the number of

channels, since the physical memory is equally

divided according to the number of channels.

On a 4-channel powermeter, the maximum is

16383 points.

On a 2-channel powermeter, the maximum is

32767 points.

On a 1-channel powermeter, the maximum is

65534 points.

The sufix [1..n], appearing after the keyword

"TRACe", is not used.

Syntax

:TRACe[1..n]:POINtsTRC1|TRC2|TRC3|T

RC4[,]

Parameter(s)

³

Trace:

The program data syntax for the first parameter is

defined as a

element. The allowed

DATA> elements for this parameter are: