Typical Applications

High Speed Power Meter

99

Testing Instrument Linearity

To confirm the linearity of the attenuator being used, you need a stable
source (IQS-2100), a variable attenuator, your high-speed power meter,
two test jumpers, and appropriate connector adapters.

Once you are confident with both the test setup and instruments being
used, you can start testing for component linearity.

To confirm the linearity of the attenuator:
1.
Perform an offset nulling if needed.

2. Make sure that the source is switched off.

3. Connect the instruments as shown below.

4. Switch on the source and allow it to stabilize (the recommended time

is 60 minutes).

5. With all instruments set to the same wavelength and the attenuator set

to minimum attenuation, take an IQS-1600 reference measurement
and switch to relative display. Next, switch the attenuator to Relative
display mode.

IQS-3100 Variable

Attenuator

IQS-1600

Power Meter

IQS-2100

Light Source

Light Source

IQS-2100

ACTIVE

Variable

IQS-3100

Attenuator

IN

OUT

Power Meter

IQS-1600

ANALOG OUT

EXT.

TRIG