SCPI Command Reference

174

IQS-1600

Product-Specific Commands—Description

:SENSe[1..n]:FREQuency:NCONtinuous

Description

This command sets the single acquisition rate in

Hz.

Syntax

:SENSe[1..n]:FREQuency:NCONtinuous<

SingleRate[HZ]>

Parameter(s)

SingleRate:
The program data syntax for is

defined as a

DATA> element followed by an optional

element. The

allowed element is

HZ.

The parameter is the Single

acquisition rate.

Example(s)

SENS:FREQ:NCON 512

Notes

This command is not executed if a data

acquisition is in progress. In that case, the

"acquisition in progress" message will be

returned.

See Also

SENSe:FREQuency:CONTinuous

SENSe:FREQuency:CONTinuous?

SENSe:FREQuency:CONTinuous:CATalog?

SENSe:FREQuency:NCONtinuous?

SENSe:FREQuency:NCONtinuous:CATalog?